EasyManua.ls Logo

Flowserve SEM-100 - Page 19

Flowserve SEM-100
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Flowserve.com 19 PUIOM000650-00-UIM (EN/HL) August 2022
SEM-100 and SEM-100FT User Instructions
7. Troubleshooting guide
Problem: Meter reads high or low.
Solution: Check for the following:
Ensure lines are full and there is no trapped air in the system
before calibrating.
Pump may be sucking air due to low uid level or bad pump
inlet connection.
Calibrating at one ow rate and operating at a different rate.
Calibrating with a different uid than what is being pumped.
Problem: Meter ashes “ERR” at the end of calibration.
Solution: Check for the following:
Meter electronics are not sensing the rotation of the magnet.
During calibration, if the CAL-1 or CAL-2 text does not blink
while uid is being pumped, rotation is not being sensed.
Disassemble meter and clean out or replace the meter
chamber (SEM-10 KIT)
Volume dispensed during calibration is too small or too large.
The meter must be calibrated with 9.5 liters – 265 liters
(2.5 gallons – 70 gallons)

Related product manuals