True-rms Multimeter
Performance Tests
9
9.
R
Continuity
600
Ω 20 Ω Beeper On
10.
600 Ω 250 Ω Beeper Off
11.
AC Volts
6.000 V 5 V, 45 Hz 4.947 to 5.053
12.
6.000 V 5 V, 1 Hz 4.897 to 5.103
13.
60.00 V 50 V, 45 Hz 49.47 to 50.53
14.
60.00 V 50 V, 1 Hz 48.97 to 51.03
15.
600.0 V 600 V, 45 Hz 593.7 to 606.3
16.
600.0 V 600 V, 1 Hz 587.7 to 612.3
17.
e
AC Volts + Hz
6.000 V 2 V, 50 kHz
[1]
NA 49.93 to 50.07
18.
DC Volts
6.000 V 0 V -0.002 to 0.002
19.
6.000 V 5 V 4.973 to 5.027
20.
60.00 V 50 V 49.73 to 50.27
21.
600.0 V 600 V 596.8 to 603.2
22.
600.0 V -600 V -596.8 to -603.2
23.
m
AC Millivolts
600.0 mV 6 mV, 45 Hz 5.6 to 6.4
24.
600.0 mV 600 mV, 1 kHz 587.7 to 612.3
25.
DC Millivolts
600.0 mV 10 mV 5.6 to 6.4
26.
600.0 mV 600 mV 587.7 to 612.3
27.
Diode 2.000 V 1.9 V NA 1.881 to 1.919
28.
Capacitance
1000 nF Open NA 0 to 2
29.
9999 μF 900 μF NA 881 to 919
30.
LoZ
Capacitance
10.00 μF9 μF NA 7.18 to 10.82
Table 3. DMM Performance Tests: 110/114/115/116/117 (cont.)
Step Function Range Applied
Display Reading
[1]
110 114 115 116 117

Other manuals for Fluke 110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 110 and is the answer not in the manual?

Fluke 110 Specifications

General IconGeneral
Product colorBrushed steel

Related product manuals