True-rms Multimeter
Performance Tests
11
Table 4. DMM Performance Tests: 113
Step Function Range Applied Display Reading
1.
e
Ohms
600.0 0.0 Ω 0.0 to 0.2 (2-Wire comp)
2.
600.0 500 Ω 495.3 to 504.7
3.
6.000 k 5 kΩ 4.954 to 5.046
4.
60.00 k 50 kΩ 49.54 to 50.46
5.
k CHEK
Continuity
NA 20 Ω Beeper On
6.
NA 250 Ω Beeper Off
7.
k CHEK Diode Test NA 1.9 V 1.859 to 1.941
8.
k CHEK Volts
[1]
6.000 V 0.1 V 0.095 to 0.105
9.
6.000 V 5 V 4.897 to 5.103
10.
6.000 V -5 V -5.103 to -4.897
11.
6.000 V 5 V, 45 Hz 4.897 to 5.103
12.
6.000 V 5 V, 1 kHz 4.797 to 5.203
13.
60.00 V 50 V 48.97 to 51.03
14.
60.00 V -50 V -51.03 to -48.97
15.
60.00 V 50 V, 500 Hz 48.97 to 51.03
16.
60.00 V 50 V, 1 kHz 47.97 to 52.03
17.
600.0 V
[2]
600 V 587.7 to 612.3
18.
600.0 V
[2]
-600 V -612.3 to -587.7
19.
600.0 V
[2]
600 V, 45 Hz 587.7 to 612.3
20.
600.0 V
[2]
600 V, 1 kHz 575.7 to 624.3
21.
Capacitance
1000 nF Open 0 to 2
22.
9999 μF 900 μF 881 to 919
[1] Manually select the range by pressing q.
[2] To keep from tripping the calibrator to standby, ramp up the voltage in 50 V increments with a five-second delay between increments.