Calibration Information
10
7. After each input value is applied, press g to accept the value and proceed to the next step (C/02 and
so forth).
Note
After pressing g, wait until the step number advances before changing the calibrator
source or turning the Meter’s rotary knob. Some adjustment steps can take up to several
seconds to execute before moving to the next step.
If the knob is not in the correct position for a given step, the meter will flash the unit
annunciators until the knob is put in a valid position. The keys that show the reading and
required input values are not allowed until the knob is correct.
Likewise, if the rotary switch is not in the correct position or the measured value is not within
the anticipated range of the input value, the Meter will emit a double beep and will not
continue to the next step when g is pressed.
8. After the final step, the display shows “End” to indicate that the calibration adjustment is complete.
Press g to return to meter mode.
Note
Set the calibrator to Standby prior to changing the function switch position and after
completing adjustment of each function.
If the calibration adjustment procedure is not properly completed, the Meter will not operate
correctly.
Table 4. Calibration Adjustment Steps
Rotary
Switch
Position
Calibration Steps
Input
Terminals
Calibrator
Source Value
C/01
+ and COM 0 V, 0 Hz
C/02
+ and COM 300 mV, 0 Hz
C/03
+ and COM 100 mV, 0 Hz
C/04
+ and COM -300 mV, 0 Hz
C/05
+ and COM 60 mV, 0 Hz
C/06
+ and COM 600 mV, 0 Hz
k CHEK
mV ac/dc
C/07
+ and COM 600 mV, 60 Hz
C/08
+ and COM 600 e, 2-wire comp
C/09
+ and COM 6 ke
C/10
+ and COM 60 ke
e
Ohms
C/11
+ and COM 600 ke
C/12
+ and COM 6 V, 60 Hz
C/13
+ and COM 60 V, 60 Hz
k CHEK
V ac
C/14
+ and COM 600 V, 60 Hz
[1]
[1] To keep from tripping the calibrator to standby, ramp up the voltage in 50 V increments with a 5 second delay between increments.