113
Calibration Information
4
Accuracy Specifications
Accuracy
± ([% of Reading] + [Counts])
Function Range Resolution
DC, 45 to 500 Hz 500 Hz to 1 kHz
k Chek
[1,2]
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
2.0 % + 3 4.0 % + 3
Function Range Resolution Accuracy
s
[3]
600 Ω 1 Ω
Beeper on <20 Ω, off >250 Ω; detects opens or
shorts of 500 μs or longer.
Ω
[3]
600.0 Ω
6.000 kΩ
60.00 kΩ
0.1 Ω
0.001 kΩ
0.01 kΩ
0.9 % + 2
0.9 % + 1
0.9 % + 1
R
[3]
2.000 V 0.001 V 2.0 % + 3
E
[3]
1000 nF
10.00 μF
100.0 μF
9999 μF
1 nF
0.01 μF
0.1 μF
1 μF
1.9 % + 2
1.9 % + 2
1.9 % + 2
100 μF - 1000 μF: 1.9 % + 2
>1000 μF: 5 % + 20
[1] All k Chek voltage ranges are specified from 60 counts to 100 % of range. Because inputs below 60 counts are not specified, it is possible and
normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit or are shorted
together.
[2] Crest factor of ≤3 at 4000 counts, decreasing linearly to 1.5 at full scale.
[3] After measuring voltage, a wait time of 1 minute is required to maintain accuracy of ohms, capacitance, diode test, and continuity.
Input Characteristics
Function Input Impedance (Nominal)
Common Mode Rejection Ratio
(1 kΩ Unbalanced)
Normal Mode Rejection
k Chek ~3 kΩ <300 pF >60 dB at dc, 50 or 60 Hz
Open Circuit Test Voltage Full Scale Voltage Short Circuit Current
Ω
<2.7 V dc
<0.7 V dc
<350 μA
R
<2.7 V dc 2.000 V dc <1.0 mA
MIN MAX Recording Accuracy and Response Time
Specified accuracy of the measurement function ±40 counts in kChek for changes >500 ms in duration, ±12 counts in Ohms for changes
>325 ms in duration. Typical 100 ms response to 80 %. Response time not specified for Capacitance.