113 
Calibration Information
 
8 
The following sections present the features and Meter pushbutton functions available during the Calibration 
Adjustment Procedure. Should the Meter fail any of the performance tests, perform the Calibration Adjustment 
Procedure. 
Use the following steps to view the Meter’s calibration counter. 
1.  While pressing q, turn the rotary switch from OFF to Ω function. The Meter should display “ZCAL”. 
2.  Press g once to view the calibration counter. For example, “n001” 
3.  Turn the rotary switch to OFF. 
Calibration Adjustment Password 
To start the Calibration Adjustment Procedure, the correct 4-digit password must be entered. The default password is 
“1234”. The password can be changed or reset to the default as described in following paragraphs.  
Changing the Password 
Use the following steps to change the Meter’s password: 
1.  While pressing q, turn the rotary switch from OFF to Ω function. The Meter should display “ZCAL”. 
2.  Press g once to see the calibration counter.  
3.  Press g again to start the password entry. The Meter displays “?.?.?.?”. 
4.  The Meter buttons indicated below represent the numbers 1 through 5 when entering or changing the 
password: 
K = 1  M = 2  q = 3  g = 4  Q = 5 
 
5.  Press 4 buttons to enter the current password. If changing the password for the first time, enter 
K (1), M (2), q (3), and g (4). 
6.  Press qto change the password. The Meter displays “////” if the entered password is correct. If the 
password is not correct, the Meter emits a double beep, displays “
?.?.?.?”, and the password must be 
entered again. Repeat step 5. 
7.  Press the 4 buttons of the new password. 
8.  Press g to store the new password. 
Restoring the Default Password 
If the calibration password is forgotten, the default password (1234) can be manually restored using the following 
steps: 
XW Warning 
To avoid electric shock or personal injury, remove the test leads and any input 
signal before removing the Meter’s back case. 
1.  Remove the Meter’s back case. Leave the pca in the top case. 
2.  Apply 9.0 V across the battery contacts (XBT1) + and (XBT2) – on the back of the PCA. See Figure 3. 
3.  Turn the rotary switch from OFF to any on position. 
4.  Short across the S7 CAL keypad on the back of the PCA. See Figure 3. The Meter should beep. The 
default password is now restored. 
5.  Remove the 9.0 V supply and replace the Meter’s back case.