12E 
Users Manual 
8 
Testing Diodes 
 Warning 
To avoid electrical shock or damage to the 
Meter when testing diodes in a circuit, make 
sure the power to the circuit is turned off and 
all capacitors are discharged. 
1.  Turn the rotary switch to 
. 
2.  Press the YELLOW function button once to activate 
Diode Test. 
3.  Connect the red test lead to the 
 terminal and the 
black test lead to the COM terminal. 
4.  Connect the red probe to the anode side and the 
black test lead to the cathode side of the diode being 
tested. 
5.  Read the forward bias voltage value on the display. 
6.  If the polarity of the test leads is reversed with diode 
polarity, the display reading shows . This can be 
used to distinguish the anode and cathode sides of a 
diode. 
Measuring Capacitance 
 Caution 
To avoid damage to the Meter, disconnect 
circuit power and discharge all high-voltage 
capacitors before measuring capacitance. 
1.  Turn the rotary switch to 
. 
2.  Connect the red test lead to the  terminal and the 
black test lead to the COM terminal. 
3.  Touch the probes to the capacitor leads. 
4.  After allowing the reading to stabilize (up to 15 
seconds), read the capacitance value on the display. 
Maintenance 
Beyond replacing batteries and fuses, do not attempt to 
repair or service your Meter unless you are qualified to do 
so and have the relevant calibration, performance test, 
and service instructions. The recommended calibration 
cycle is 12 months.