Multimeter
Accuracy Specifications
13
Function Range Resolution Accuracy
Resistance (Ohms)
400.0 Ω
4.000 kΩ
40.00 kΩ
400.0 kΩ
4.000 MΩ
40.00 MΩ
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.5% + 3
0.5% + 2
0.5% + 2
0.5% + 2
0.5% + 2
1.5% + 3
Capacitance
3
50.00 nF
500.0 nF
5.000 μF
50.00 μF
100.0 μF
0.01 nF
0.1 nF
0.001 μF
0.01 μF
0.1 μF
2% + 5
2% + 5
5% + 5
5% + 5
5% + 5
AC Current (40 to 200 Hz)

400.0 μA
4000 μA
0.1 μA
1 μA
1.5% + 3
AC Current (40 to 200 Hz)

40.00 mA
400.0 mA
0.01 mA
0.1 mA
1.5% + 3
AC Current (40 to 200 Hz)
4.000 A
4
10.00 A
0.01 A
0.01 A
1.5% + 3
DC Current

400.0 μA
4000 μA
0.1 μA
1 μA
1.5% + 3
3. Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the 50 nF range).
4. When in the 4A range, display will show 4000 counts, please ignore the last digit.