EasyManua.ls Logo

Fluke 1508 - Earth-Bond Resistance Measurement Specifications; Insulation Measurement Specifications

Fluke 1508
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Insulation Testers
Specifications
23
Earth-Bond Resistance Measurement
Range Resolution
Accuracy
1
+
(% of Rdg + Digits)
20.00 Ω 0.01 Ω
200.0 Ω 0.1 Ω
2000 Ω 1 Ω
20.00 kΩ 0.01 kΩ
+
(1.5 % + 3)
1. Accuracies apply from 0 to 100% of range.
Overload Protection ............................................2 V rms or dc
Open Circuit Test Voltage ...................................> 4.0 V, < 8 V
Short Circuit Current............................................> 200.0 mA
Insulation Specifications
Measurement Range...........................................0.01 MΩ to 10 GΩ
Test Voltages ......................................................50, 100, 250, 500, 1000 V
Test Voltage Accuracy ........................................+ 20 %, - 0 %
Short-Circuit Test Current ...................................1 mA nominal
Auto Discharge....................................................Discharge time <0.5 second for C = 1 μF or less
Live Circuit Detection: .........................................Inhibit test if terminal voltage > 30 V prior to initialization of test.
Maximum Capacitive Load..................................Operable with up to 1 μF load.

Other manuals for Fluke 1508

Related product manuals