Users Manual
14
The Meter beeps when the reading is stable. The upper right display shows the same
resistance reading as the main display.
When resistance is higher than the maximum display range, the meter reacts as
follows:
If the 250V range is selected, the display reads >1000 M.
If the 500V range, is selected, the display reads >2000 M.
If the 1000V range, is selected, the display reads >4000 M.
4. While keeping the probes on the test points, release the TEST button.
The circuit now discharges through the Meter, while the main reading shows the
decreasing voltage.
Keep the probes touched to the test points until the circuit is completely discharged
(the main display shows - - - -).
The upper right display holds the resistance reading until a new test is started or a
different function is selected.