1535/1537
Calibration Manual
12
adc i r Read the Current analog-to-digital conversion value ADC_READ “current_adc”
cal_param current 7
-19 ADC_READ
Write the calibration parameter 7, use read value
ADC_READ in the previous step
“current param”
irs 1 Enter 200 μA range “irs”
Current source outputs -25 μA
adc i r Read the Current analog-to-digital conversion value ADC_READ “current_adc”
cal_param current 8
-25 ADC_READ
Write the calibration parameter 8, use read value
ADC_READ in the previous step
“current param”
Current source outputs -190 μA
adc i r Read the Current analog-to-digital conversion value ADC_READ “current_adc”
cal_param current 9
-190 ADC_READ
Write the calibration parameter 9, use read value
ADC_READ in the previous step
“current param”
irs 0 Enter 2 mA range “irs”
Current source outputs -250 μA
adc i r Read the Current analog-to-digital conversion value ADC_READ “current_adc”
cal_param current 10
-250 ADC_READ
Write the calibration parameter 10, use read value
ADC_READ in the previous step
“current param”
Current source outputs -400 μA
adc i r Read the Current analog-to-digital conversion value ADC_READ “current_adc”
cal_param current 11
-400 ADC_READ
Write the calibration parameter 11, use read value
ADC_READ in the previous step
“current param”
cal stop_current Stop the Current calibration “cal stop_current”
cal_param save Save the calibration parameters “params saved”