1535/1537
Calibration Manual
2
Safety Information
A Warning identifies hazardous conditions and procedures that are dangerous to the user. A Caution
identifies conditions and procedures that can cause damage to the Product or the equipment under test.
XW Warning
To prevent possible electrical shock, fire, or personal injury:
Carefully read all instructions.
Read all safety information before you use the Product.
Do not alter the Product and use only as specified, or the protection supplied by
the Product can be compromised.
Do not use the Product around explosive gas, vapor, or in damp or wet
environments.
Do not use the Product if it is altered or damaged.
Disable the Product if it is damaged.
Do not use the Product if it operates incorrectly.
Use Product-approved measurement category (CAT), voltage, and amperage
rated accessories (probes, test leads, and adapters) for all measurements.
Do not exceed the Measurement Category (CAT) rating of the lowest rated
individual component of a Product, probe, or accessory.
Do not use in CAT III or CAT IV environments without the protective cap installed
on test probe. The protective cap decreases the exposed probe metal to <4 mm.
This decreases the possibility of arc flash from short circuits.
Comply with local and national safety codes. Use personal protective
equipment (approved rubber gloves, face protection, and flame-resistant
clothes) to prevent shock and arc blast injury where hazardous live conductors
are exposed.
Examine the case before you use the Product. Look for cracks or missing
plastic. Carefully look at the insulation around the terminals.
Do not use test leads if they are damaged. Examine the test leads for damaged
insulation, exposed metal, or if the wear indicator shows. Check test lead
continuity.
Do not touch voltages >30 V ac rms, 42 V ac peak, or 60 V dc.
Do not apply more than the rated voltage, between the terminals or between
each terminal and earth ground.
Measure a known voltage first to make sure that the Product operates correctly.
Limit operation to the specified measurement category, voltage, or amperage
ratings.
The battery door must be closed and locked before you operate the Product.
Connect the common test lead before the live test lead and remove the live test
lead before the common test lead.
Remove all probes, test leads, and accessories before the battery door is
opened.
Remove all probes, test leads, and accessories that are not necessary for the
measurement.
Keep fingers behind the finger guards on the probes.
Replace the batteries when the low battery indicator shows to prevent incorrect
measurements.
Use the correct terminals, function, and range for measurements.