Insulation Tester
Safety Information
3
Place test leads in proper input terminals.
Do not work alone.
Do not use in distribution systems with voltages higher than 660 V.
Use only recommended test leads.
Remove all power from the circuit under test and discharge circuit capacitance
before testing resistance or capacitor with the tester.
Results of measurement can be adversely affected by the impedances of
additional operating circuits connected in parallel or by transient currents.
Before and after testing, confirm that the Tester does not indicate the presence
of a hazardous voltage. If a hazardous voltage is shown on the display, remove
power from the circuit under test or allow the installation capacitance to fully
discharge.
Do not disconnect the test leads before a test has been completed and the test
voltage at the terminals has returned to zero. This ensures that any charged
capacitance is fully discharged.
Use the guard terminal only as specified in this manual. Do not allow other
foreign objects coming into contact with the guard terminals as safety may be
compromised
Remove the input signals before you clean the Product.
Use only specified replacement parts.
Repair the Product before use if the battery leaks.
Be sure that the battery polarity is correct to prevent battery leakage.
Do not operate the Product with covers removed or the case open. Hazardous
voltage exposure is possible.
Have an approved technician repair the Product.