Digital Multimeter
Accuracy Specifications
5
Accuracy Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to
28 °C, relative humidity at 0 % to 75 %.
Accuracy specifications take the form of: ±([% of Reading] + [Number of Least
Significant Digits])
Accuracy
Function Range Resolution
Model 15B Model 17B
400.0 mV
1
0.1 mV 3.0 % + 3 3.0 % + 3
AC Volts (40 to 500 Hz)
v
4.000 V
40.00 V
400.0 V
1000 V
0.001 V
0.01 V
0.1 V
1 V
1.0 % + 3 1.0 % + 3
DC Millivolts
M
400.0 mV 0.1 mV 1.0 % + 10 1.0 % + 10
DC Volts
V
4.000 V
40.00 V
400.0 V
1000 V
0.001 V
0.01 V
0.1 V
1 V
0.5 % + 3 0.5 % + 3
Diode Test
2
1.000 V 0.001 V 10 %
Temperature
3
°C
(K-type thermocouple)
50 °C to 400 °C
0 °C to 50 °C
-55 °C to 0 °C
0.1 °CNA 2 % + 1 °C
± 2 °C
9 % + 2 °C
Resistance (Ohms)
400.0
4.000 k
40.00 k
400.0 k
4.000 M
40.00 M
0.1
0.001 k
0.01 k
0.1 k
0.001 M
0.01 M
0.5 % + 3
0.5 % + 2
0.5 % + 2
0.5 % + 2
0.5 % + 2
1.5 % + 3
0.5 % + 3
0.5 % + 2
0.5 % + 2
0.5 % + 2
0.5 % + 2
1.5 % + 3
Capacitance
4
c
50.00 nF
500.0 nF
5.000 µF
50.00 µF
100.0 µF
0.01 nF
0.1 nF
0.001 µF
0.01 µF
0.1 µF
2 % + 5
2 % + 5
5 % + 5
5 % + 5
5 % + 5
2 % + 5
2 % + 5
5 % + 5
5 % + 5
5 % + 5
Frequency
Hz
(10 Hz 100 kHz)
50.00 Hz
500.0 Hz
5.000 kHz
50.00 kHz
100.0 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz
NA 0.1 % + 3
Duty Cycle 0.1 % to 99.9 % 0.1% NA 1% typical
5
1. Manual Range only
2.
Diode test open circuit test voltage is 1.1 V to 1.6 V and short circuit current is < 0.6 mA (typical).
3.
Temperature specifications do not include thermocouple errors. After inserting the thermocouple plug into the meter, allow several
minutes for thermal stabilization.
4.
Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the 50 nF range).
For the 17B, errors can be reduced by using the relative feature.
5.
For values between 10% and 90% duty cycle at 50 Hz.

Other manuals for Fluke 15B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 15B and is the answer not in the manual?

Fluke 15B Specifications

General IconGeneral
ContinuityYes
Diode TestYes
Battery2 x AA
Safety RatingCAT III 600V
Display4, 000 count digital display