EasyManuals Logo

Fluke 15B Calibration Manual

Fluke 15B
30 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
Digital Multimeter
Meter Adjustment
15
Meter Adjustment
If the Meter fails the performance tests, adjustment may be necessary. Use the steps listed
below and refer to Table 5 or 6 to adjust the Meter.
1. Remove the yellow holster by slipping it off of the Meter.
2. Remove the screws on the bottom of the Meter.
3. Lift off the bottom case to gain access to the potentiometers used for Meter
adjustment. Refer to Figure 6.
4. Observing correct polarity, connect a 3 V dc lab supply to the meters battery
contacts to power the meter during calibration. Refer to Figure 5 for proper
connections.
5. Apply the values listed in Table 5 or 6 and adjust the designated potentiometer as
needed.
6. When finished, replace the bottom case.
7. Replace the case screws.
8. Replace the yellow holster.
9. Verify the meter adjustment by going through the performance test procedures. If the
Meter still does not meet the performance test specifications, service is required.
Refer to Contacting Fluke to locate a service center.

Other manuals for Fluke 15B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 15B and is the answer not in the manual?

Fluke 15B Specifications

General IconGeneral
ContinuityYes
Diode TestYes
Battery2 x AA
Safety RatingCAT III 600V
Display4, 000 count digital display

Related product manuals