Digital Multimeter
Performance Tests
11
Table 4. 17B Performance Test
Step
Test
Function
UUT Switch
Setting
UUT Push
Button
5520A Output UUT Display limits
1 3.5 V 50 Hz 3.462 to 3.538 V ac
2 35 V 500 Hz 34.62 to 35.38 V ac
3
N/A
350 V 50 Hz 346.2 to 353.8 V ac
4 R
5 1000 V 50 Hz Auto Range icon
OFF
OL.
6 R 987 to 1013 V ac
7 R 0.35 V 50 Hz 339.2 to 360.8 mV ac
8 I 1 V 50Hz 49.92 to 50.08 Hz
9 - 1 V 500 Hz 499.2 to 500.8 Hz
10 - 1 V 5 kHz 4.992 to 5.008 Hz
11 - 2 V 50 kHz 49.92 to 50.08 Hz
12 - 3 V 100kHz 99.6 to 100.4 Hz
13
Volts AC
v
I 3 V 500 Hz 49.0 to 51.0 %