15B & 17B
Calibration Manual
10
Table 3. 15B Performance Test (cont.)
Step
Test
Function
Switch
Position
Push Button
5520A
Output
UUT Reading
Limits
For the 15B Capacitance Test, subtract the test lead capacitance from the UUT reading to get
measurement result.
22 48 nF 46.99 to 49.01 nF
23 0.48 µF 469.9 to 490.1 µF
24
Capacitance N/A
4.8 µF 4.555 to 5.045 µF
For step 25, wait 15 seconds for the reading to settle.
25 - - - 48 µF 45.55 to 50.45 µF
For step 26, wait 15 seconds for reading to settle.
26 - - - 95 µF 89.7 to 100.3 µF
27 Amps dc A - 3.8 A dc 3.713 to 3.887 A dc
28 - - - -10 A dc -9.82 to -10.18 A dc
29 Amps ac - C 3.8 A 50 Hz 3.713 to 3.887 A ac
30 - - - 10 A, 50 Hz 9.82 to 10.18 A ac
31 mA dc m - 38 mA dc 37.40 to 38.60 mA dc
32 - - - 0.38 A dc 374.0 to 386.0 mA dc
33 mA ac - C 0.38 A 50 Hz 374.0 to 386.0 mA ac
34 - - - 38 mA 200 Hz 37.40 to 38.60 mA ac
35 µA dc N - 3.8 mA dc 3740 to 3860 µA dc
36 - - - 0.38 mA dc 374.0 to 386.0 µA dc
37 µA ac - C 0.38 mA, 40 Hz 374.0 to 386.0 µA ac
38 - - - 3.8 mA, 200 Hz 3740 to 3860 µA ac
The Fluke 15B Performance Test is now complete.