Fluke 15B Calibration Manual

Fluke 15B
30 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
Digital Multimeter
Performance Tests
9
Table 3. 15B Performance Test
Step
Test
Function
Switch
Position
Push Button
5520A
Output
UUT Reading
Limits
1 3.5 V 50 Hz 3.462 to 3.538 V ac
2 35 V 500 Hz 34.62 to 35.38 V ac
3
N/A
350 V 50 Hz 346.2 to 353.8 V ac
4
R
5 1000 V 50 Hz Auto Range
icon OFF
OL.
6
R
987 to 1013 V ac
7
Volts AC
v
R
0.35 V 60 Hz 339.2 to 360.8 mV ac
Set 5520A to Standby
8 3.5 V dc 3.480 to 3.520 V dc
9 35 V dc 34.80 to 35.20 V dc
10 350 V dc 348.0 to 352.0 V dc
11
Volts DC N/A
-1000 V dc -992 to -1008 V dc
Set 5520A to Standby
12 mV dc M N/A 0.35 V dc 345.5 to 354.5 mV dc
13 350 347.9 to 352.1
14 3.5 k 3.480 to 3.520 k
15 35 k 34.80 to 35.20 k
16 0.35 M 348.0 to 352.0 M
17 3.5 M 3.480 to 3.520 M
18
Ohms
F N/A
35 M 34.45 to 35.55 M
19 Diode F C
(once)
0.7 V dc D icon ON
0.630 to 0.770 V dc
20 Continuity F C
(press again)
40 E Icon ON
Beeper ON
21 - - - 800 Beeper OFF

Other manuals for Fluke 15B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 15B and is the answer not in the manual?

Fluke 15B Specifications

General IconGeneral
ContinuityYes
Diode TestYes
Battery2 x AA
Safety RatingCAT III 600V
Display4, 000 count digital display