EasyManua.ls Logo

Fluke 1625 - AC Resistance Measurement Specs (R~)

Fluke 1625
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1625
Users Manual
18
Minimal current in single
branch to be measured:
0.5 mA with transformer (1000:1)
0.1 mA with transformer (200:1)
Max. interference current
through transformer:
3 A with a transformer (1000:1)
Resistance Measurement (R~)
Measuring method: current and voltage measurement
Measuring voltage: 20 V AC, square pulse
Short circuit current: > 250 mA AC
Measuring frequency: 94, 105, 111, 128 Hz selected manually or auto-
matically (AFC)
Measuring
Range
Display Range Resolution
Intrinsic
Error
Operating
Errors
0.001 ... 2.999 0.001
3.0 ... 29.99 0.01
0.020 ... 30 ... 299.9 0.1
300 k 300 ... 2999 1
± ( 2 %
of m.v.
+2 digit )
± ( 5% of m.v.
+5 digit )
3.0 k ... 29.99 k 10
30.0 k ... 299.9 k 100
Measuring time: typ. 6 sec.
Max. interference voltage: 24 V, with higher voltages measurement will not
be started
Max overload: Urms max. = 250 V

Table of Contents

Related product manuals