Page #26 background image
1625
Users Manual
18
Minimal current in single
branch to be measured:
0.5 mA with transformer (1000:1)
0.1 mA with transformer (200:1)
Max. interference current
through transformer:
3 A with a transformer (1000:1)
Resistance Measurement (R~)
Measuring method: current and voltage measurement
Measuring voltage: 20 V AC, square pulse
Short circuit current: > 250 mA AC
Measuring frequency: 94, 105, 111, 128 Hz selected manually or auto-
matically (AFC)
Measuring
Range
Display Range Resolution
Intrinsic
Error
Operating
Errors
0.001 ... 2.999 0.001
3.0 ... 29.99 0.01
0.020 ... 30 ... 299.9 0.1
300 k 300 ... 2999 1
± ( 2 %
of m.v.
+2 digit )
± ( 5% of m.v.
+5 digit )
3.0 k ... 29.99 k 10
30.0 k ... 299.9 k 100
Measuring time: typ. 6 sec.
Max. interference voltage: 24 V, with higher voltages measurement will not
be started
Max overload: Urms max. = 250 V

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1625 and is the answer not in the manual?

Fluke 1625 Specifications

General IconGeneral
Storage Temperature-20 °C to 60 °C
Testing Methods3-pole fall-of-potential, 4-pole
Measurement Frequency128 Hz
Power Supply6 x 1.5 V AA batteries or optional AC adapter
Safety RatingCAT II 300 V
Safety StandardsIEC 61010-1