1625
Users Manual
20
Compensation of Lead Resistance (R
K
)
Compensation of lead resistance (R
K
) can be
switched on in functions R
E
3pole, R
E
4pole
A, R~ , and R F 2pole
Formation of measured value:
R
display
= R
measured
- R
compensated
*
* Value of setpoint entry R
K
= 0.000, vari-
able from 0.000…29.99 by means of
measuring adjustment.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1625 and is the answer not in the manual?

Fluke 1625 Specifications

General IconGeneral
Storage Temperature-20 °C to 60 °C
Testing Methods3-pole fall-of-potential, 4-pole
Measurement Frequency128 Hz
Power Supply6 x 1.5 V AA batteries or optional AC adapter
Safety RatingCAT II 300 V
Safety StandardsIEC 61010-1