54
6. The changed value is stored by pressing the "DISPLAY MENU" key.
7. Pressing the "START TEST" key exits the setting program.
Note
If the limit values required by regulations are changed incorrect, er-
roneous test results may be displayed.
Deleting a Code
1. Press all 4 keys simultaneously and move central selector from OFF
to any measuring mode.
The display shows " C _ _ _ ".
2. Now enter the existing CODE-number.
3. Inputing the code is done by means of the "CHANGE ITEM" and
"SELECT" keys. Pressing the "DISPLAY MENU" key completes in-
put.
4. Display shows "C ON". In the "C ON" state the CODE function can
be disabled by pressing the "CHANGE ITEM" key. The display then
shows "C OFF".
5. If this display is acknowledged by pressing the "DISPLAY MENU"
key, the user code and all changes of the limit values are erased. The
original default values are restored into memory.
6. Now a new CODE-number may be programmed and used for setting
new parameters.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1625 and is the answer not in the manual?

Fluke 1625 Specifications

General IconGeneral
Storage Temperature-20 °C to 60 °C
Testing Methods3-pole fall-of-potential, 4-pole
Measurement Frequency128 Hz
Power Supply6 x 1.5 V AA batteries or optional AC adapter
Safety RatingCAT II 300 V
Safety StandardsIEC 61010-1

Related product manuals