1773/1775/1777
Calibration Manual
22
17xx Auxiliary Input Adapter Verification
The Auxiliary Input Adapter has a 1000:1 divider that can be verified with a calibrator and an
8588A. To connect to the pins, use a banana-to-pin adapter. Pomona Electronics 4690 is
recommended. See Ta bl e 1 5 and Ta bl e 1 6 .
Table 15. 17xx AUX Adapter Pin-out
Pin Signal
1AUX 1 +
2AUX 1 -
3AUX 2 +
4AUX 2 -
Table 16. 17xx AUX Voltage Divider Input
Input Range
Intrinsic Accuracy AUX Adapter + Instrument
(% of Reading + % of Range)
Direct Input ±10 V see instrument specification
Voltage divider input ±1000 V ±(0.5 % + 0.002 V)
Note: Environmental Reference Conditions: 23 °C ±5 °C, instrument operating for at least
30 minutes, no external electrical/magnetic field, RH <65 %.