EasyManuals Logo

Fluke 17B MAX Calibration Manual

Fluke 17B MAX
14 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #10 background imageLoading...
Page #10 background image
15B MAX/17B MAX
Calibration Manual
6
Step Function
Calibrator Output
15B
MAX
17B
MAX
Lower
Limit
Upper
Limit
Value
Frequency or
Amplitude
16
OHM
0 ΩΩ √√ -0.3 0.3
17 350 ΩΩ √√ 348.0 352.0
18 3.5 kΩΩ √√ 3.480 3.520
19 35 kΩΩ √√ 34.80 35.20
20 350 kΩΩ √√ 348.0 352.0
21 3.5 MΩΩ √√ 3.480 3.520
22 10 MΩΩ √√ 9.82 10.18
23 BEEPER ON 40 ΩΩ √√ XX
24 DIODE 0.7 V 0 Hz √√ 0.622 0.778
25
CAP
35 nF Ω √√ 34.24 35.76
26 350 nF Ω √√ 342.4 357.6
27 3.5 μF Ω √√ 3.320 3.680
28 35 μF Ω √√ 33.20 36.80
29 350 μF Ω √√ 332.0 368.0
30
ADC
3.5 A 0 Hz √√ 3.444 3.556
31 -10 A 0 Hz √√ -10.18 -9.82
32
AAC
3.5 A 50 Hz √√ 3.444 3.556
33 10 A 400 Hz √√ 9.82 10.18
34
mADC
35 mA 0 Hz √√ 34.44 35.56
35 -350 mA 0 Hz √√ -355.6 -344.4
36
mAAC
35 mA 400 Hz √√ 34.44 35.56
37 350 mA 50 Hz √√ 344.4 355.6
38
μADC
350 μA0 Hz √√ 344.4 355.6
39 -3500 μA0 Hz √√-3556 -3444
40
μAAC
350 μA 40 Hz √√ 344.4 355.6
41 3500 μA 400 Hz √√ 3444 3556
42
Te m p
-50 °C X X -56.4 -43.6
43 35 °C X X 33.0 37.0
44 400 °C X X 391.0 409.0
Table 2. Performance Specifications (cont.)

Other manuals for Fluke 17B MAX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 17B MAX and is the answer not in the manual?

Fluke 17B MAX Specifications

General IconGeneral
BrandFluke
Model17B MAX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals