Performance Tests
5
Ta b l e 1 lists the equipment required for the performance tests.
Before you do the performance test, make sure to warm up the calibrator.
To test each function and operating ranges:
1. For each step in Ta b l e 2 , set the DUT to the specified function and range.
2. Connect the source to the input jacks on the DUT.
3. Apply the output from the source. For temperature connections, see Figure 3.
4. Make sure the DUT readings are within the limits in Ta bl e 2 .
Table 1. Required Equipment
Equipment Recommended Model
Calibrator Fluke Calibration 5502A, 5502E, or 5522A
Type K to dual banana adapter Fluke 80AK-A
Table 2. Performance Specifications
Step Function
Calibrator Output
15B
MAX
17B
MAX
Lower
Limit
Upper
Limit
Value
Frequency or
Amplitude
1
VAC
5.5 V 50 Hz √√ 5.442 5.558
2 55 V 50 Hz √√ 54.42 55.58
3 55 V 500 Hz √√ 54.42 55.58
4 550 V 50 Hz √√ 544.2 555.8
5 1000 V 50 Hz √√ 986 1014
6
VAC/Freq.
45 Hz 1 V X 44.92 45.08
75 kHz1 VX 4.992 5.008
8 100 kHz 3 V X 99.6 100.4
9 VAC/Dut y 50 Hz
3 V (Square
Wave)
X 49.5 50.5
10
VDC
5.5 V 0 Hz √√ 5.470 5.531
11 55 V 0 Hz √√ 54.70 55.31
12 550 V 0 Hz √√ 547.0 353.1
13 -1000 V 0 Hz √√-1008 -992
14 mVDC 550 mV 0 Hz √√ 543.5 556.5
15 mVAC 550 mV 500 Hz √√ 533.2 566.8