Adjustment Procedures
7
Figure 3. Temperature Connections
Adjustment Procedures
Ta b l e 3 lists the equipment required to adjust the Product.
To enter adjustment mode:
1. Remove the battery door and batteries. See Figure 2.
2. Remove the calibration sticker. See Figure 4.
Table 3. Required Equipment for Adjustment
Equipment Required Characteristics Recommended Model
Calibrator - Fluke 5522A
Potentiometer (17B MAX) Adjust Manually
Type K to dual banana (m)
adapter
- Fluke 80AK-A