Model 16 English Instruction Sheet
Page
15
MIN MAX Recording with Elapsed Time
Elapsed Time Resolution Accuracy
0 to 100 hours (99:59) 1 minute 0.3% typical
Maximum Voltage
Between any Terminal
and Earth Ground:
600V rms
Display: 3 3/4-digits, 4000 counts, updates 4/sec
Operating Temperature: -10°C to 50°C (14°F to 122°F)
Storage Temperature:
-30°C to 60°C (-22°F to 140°F)
indefinitely (to -40°C (-40°F) for 100 hrs)
Temperature
Coefficient:
(.1 x specified accuracy)/°C (<18°C or
>28°C)
Relative Humidity: 0% to 90% (-10°C to 35°C; 14°F to 95°F)
0% to 70% (35°C to 50°C; 95°F to 122°F)
Battery Type: 9V, NEDA 1604 or IEC 6F22
Battery Life: 650 continuous hours with alkaline
450 continuous hours with carbon-zinc
Shock: 1 meter drop 6 sides.
Vibration: Class II vibration MIL-PRF-28800F
Size (H x W x L): 3.46 cm x 7.05 cm x 14.23 cm
(1.35 in x 2.75 in x 5.55 in)
Weight: 286g (10 oz)
Safety:
Designed to Protection Class II
requirement of UL3111, ANSI/ISA-S82,
CSA C22.2 No 231, and VDE 0411, and
IEC 1010 overvoltage Category III (CAT
III, 600 Volts).
EMI Regulations:
Complies with FCC Part 15, Class B, and
VDE 0871B. Trademark of TÜV Product
Services. Complies with EN 61010-1:
1993.
Certifications:
P $