287/289 
Users Manual 
14 
VAC
Menu
REL
Peak, CF
Hz, %, mS
dBm
dBV
REL REL% Close
 
est05.eps 
Figure 5. Function Menu 
Menu selection is indicated by the filled-in black square 
(hereafter the menu selector) to the left of a menu item. Use the 
four front-panel cursor buttons (5 6 7 8) to position the menu 
selector next to a menu item. As the menu selector moves 
between menu items, the four softkeys and their labels change to 
reflect the available functions and/or modes available for the 
selection menu item. 
The example menu in Figure 5 shows the REL (Relative) 
function as the current selection. The function selected when the 
menu opens is the function selected the last time the menu was 
used. To get to the Hz menu item from the REL item, press 6 
once, followed by one press of 7. As the menu selector moves 
between the menu items, the softkey labels change to indicate 
each softkey’s function. Once the desired function or mode 
appears in one of the softkey labels, press the appropriate 
softkey to activate it. The pop-up menu closes and the display 
changes to reflect the selection just made. Pressing the softkey 
labeled Close closes the pop-up menu, leaving the Meter in the 
state it was in before pressing the Menu softkey. 
In most cases, the softkeys revealed by the menu selection act 
like toggles. The example menu shown in Figure 5 shows REL, 
REL%, and Close softkeys. In this example, the Meter is not in 
the relative mode, so pressing the softkey labeled REL would 
activate, or toggle, the relative mode. If, on the other hand the 
Meter is already in the relative mode, pressing the same softkey 
would disable the relative function.  
In some cases, pressing a function that can not be used with 
other functions appearing in the menu turns off the previously 
selected function. For example, in Figure 5, if the Meter is 
already in the relative function, pressing REL% causes the Meter 
to turn off relative and display relative percent.  
In cases where multiple modes have been selected, selecting the 
first (top-left) menu item always turns off all other functions and 
modes, and returns the Meter to the primary function selected by 
the rotary switch. For example, assume that the Meter is setup 
for frequency (Hz) and is displaying in relative mode as selected 
through the menu in Figure 5. Moving the menu selector to the 
menu item labeled VAC and pressing the softkey labeled VAC, 
clears both frequency and relative selections, leaving the Meter 
in volts ac only. 
Menu selections are remembered for each rotary switch position. 
For example, selecting REL for the volts ac position causes REL 
to be selected the next time the menu is opened in volts ac, even 
though in the interim, Hz,%,ms was selected from a similar menu 
for the millivolts ac function. 
Up to two columns of four items each are displayed at any one 
time. If more than eight menu items are available for a primary 
function, g appears in the lower right-hand corner of the page 
area of the display, indicating more menu items are available. 
With the menu selector on one of the items in the left column,