EasyManua.ls Logo

Fluke 360 - Safety Specifications; Electrical Specifications; AC Current Measurement; Maintenance

Fluke 360
10 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
360
AC Leakage Current Clamp Meter
4
Dimensions (W x H x D)........................................ 70 mm x 176 mm x 25 mm (2.8 in x 6.9 in x 1.0 in)
Weight (including battery).................................... Approximately 200 g (0.44 lb)
Safety Specifications
Category rating...................................................... CAT III 300 V per IEC/EN61010-1, and 61010-2-032 Pollution Degree
2, indoor use
EMC ........................................................................ EN61326-1
Effect of radiated immunity.................................. Rated accuracy + 4.0 % in an electromagnetic field of 3 v/m
Operating altitude ................................................. 2,000 m (6,562 ft)
Accessories........................................................... C75
soft carrying case
Electrical Specifications
Reference conditions ........................................... 23 ± 5 °C and 80 % RH maximum
AC Current Measurement
Range Resolution
Accuracy (50/60 Hz)
± (% of reading + digit)
Maximum Allowable Current
3 mA 0.001 mA
30 mA 0.01 mA
1 % + 5
30 A 0.01 A
60 A 0.1 A
0 to 50 A: 1 % + 5
50 to 60 A: 5 % + 5
60 A rms
Zero correction: Fractions smaller than approximately 0.01 mA are calibrated to zero.
Maintenance
Basic maintenance for the Meter includes cleaning, checking the battery charge, and replacing
depleted batteries.
Cleaning
Periodically wipe the case with a damp cloth and mild detergent.
W Caution
To avoid damage, do not use abrasives or solvents to clean the
meter.
Open the jaws and blow air to remove dust or debris at the jaw mating surfaces.
Battery Replacement
If battery voltage drops below the operating voltage, the battery symbol (M) appears on the
display. If this happens, replace the battery as soon as possible.
XW Warning
To avoid risk of electric shock during battery replacement,
always remove the Meter from measurable conductors under
test before replacing the battery. Always use the correct
replacement battery (see Table 5).
To replace the battery, complete the following steps and refer to Figure 1:
1. Ensure the power is off.
2. Place the Meter on its face and turn the battery cover counterclockwise using a
Phillips screwdriver.
3. Remove the cover and the installed battery.
4. Insert the battery ensuring the polarity is correct.
5. Close the battery cover, aligning the arrow’s tail with the dot and then turn
clockwise.

Other manuals for Fluke 360

Related product manuals