AC Leakage Current Clamp Meter
Performance Tests
5
epq003.eps
Figure 1. Changing the Battery
Performance Tests
XWWarning
To avoid possible electric shock, do not perform the
performance test procedures unless the Meter is fully
assembled and you are qualified to do so.
The following performance tests verify the complete operation of the Meter and check the
accuracy of each Meter function against its specifications. The recommended calibration interval
is 12 months. In the performance tests, the Meter is referred to as the unit under test (UUT).
Before you perform any of the following tests, check the battery and replace if necessary. For
more information, see Battery Replacement. If the UUT fails any performance test, contact Fluke
Service for repair. See Contacting Fluke.
Required Equipment
Table 2 lists the equipment required to complete the performance tests.
Table 2. Required Equipment
Equipment Minimum Required Characteristics Recommended Model
Calibrator
AC current:
• Accuracy
ACmA: ± 0.25 %
Amps: 30 A = ± 0.19%
60 A = ± 0.49%
• Frequency = 50/60 Hz
Fluke 5520A High Performance
Multi-Product Calibrator
Magnet wire coil
3 turns, 14-gauge film-coated copper wire,
6-in. diameter
---
Magnet wire coil
2 turns, 14-gauge film-coated copper wire,
6-in. diameter
---
Copper wire coil 1 turn, 14-gauge copper wire, 6-in. diameter ---