AC Leakage Current Clamp
Calibration
7
Calibration
 Warning
To avoid possible electric shock, do not perform the
performance test procedures unless the Product is fully
assembled and you are qualified to do so.
The performance tests in this section verify the complete operation of the Product
and check the accuracy of each function against its specifications. The
recommended calibration interval is 12 months. In the performance tests, the
Product is referred to as the unit under test (UUT).
Before you do any of the tests, check the battery and replace if necessary. For
more information, see Replace the Batteries.
Required Equipment
The equipment in Table 2 is necessary for the tests in this manual.
Table 2. Required Equipment
Equipment
Required
Characteristics
Recommended Model
Calibrator 4.5-digit resolution Fluke 552xA
Wired coil 50 turns 5500A/COIL
Transconductance amplifier 4.5-digit resolution Fluke 52120A
Test Lead for other PN 2070140