AC Leakage Current Clamp
Calibration
9
AAC Adjustment Procedure
For the tests where only the 552xA is used:
1. Connect the UUT as shown in Figure 3.
2. Connect the test lead to the AUX (Auxiliary Output) terminals and set LCOMP
on the 552xA to OFF.
3. Apply the test values in Table 3.
0
/
+
5522A
idh04.eps
Figure 3. AAC Connections without 50-turn Coil