368/369/368 FC/369 FC
Calibration Manual
10
For the tests where the 50-turn coil is used:
1. Connect the UUT as shown in Figure 4.
2. Connect the test lead to the AUX (Auxiliary Output) terminals, and set
LCOMP on the 552xA to ON.
3. Apply the test values in Table 3.
0
/
+
5522A
50-Turn Current Coil
idh003.eps
Figure 4. Connection with 50-turn Coil