Enter Calibration Mode
To put the Product into calibration mode:
1. Use a screwdriver to loosen the battery compartment door screw.
2. Remove the batteries.
3. The recessed calibration button is underneath the calibration sticker that is
underneath the batteries. Use a small probe to push though the sticker and
short the two sections of the calibration button traces, see
in Figure 2.
1
idh001.eps
Figure 2. Calibration Button
Calibration Process
Once the Product is in calibration mode, follow these steps to calibrate the
Product.
1. Set calibrator output to the correct signal.
2. Apply the signal to the UUT.
3. Wait 15 seconds for the reading to become stable.
4. Once stable, push to accept the reading and to move forward to the next
calibration step.
5. Push to save the calibration data.
For the tests in these sections, if the values do not meet those in Table 3,
complete the calibration and retest the Product or return it to Fluke for service.