AC Leakage Current Clamp
Calibration
13
Table 3. AAC Adjust Procedure
Step LCD display
Calibrator output
for all models
Equipment
1 C-00 0 A, 0 Hz Only use 552xA
2 C-01
30 μA, 60 Hz
Only use 552xA
3 C-02
100 μA, 60 Hz
Only use 552xA
4 C-03
300 μA, 60 Hz
Only use 552xA
5 C-04
300 μA, 60 Hz
Only use 552xA
6 C-05 0 A, 0 Hz Only use 552xA
7 C-06
300 μA, 60 Hz
Only use 552xA
8 C-07 1 mA, 60 Hz Only use 552xA
9 C-08 3 mA, 60 Hz Only use 552xA
10 C-09 3 mA, 60 Hz Only use 552xA
11 C-10 0 A, 0 Hz Only use 552xA
12 C-11 3 mA, 60 Hz Only use 552xA
13 C-12 10 mA, 60 Hz Only use 552xA
14 C-13 30 mA, 60 Hz Only use 552xA
15 C-14 30 mA, 60 Hz Only use 552xA
16 C-15 0 A, 0 Hz Only use 552xA
17 C-16 30 mA, 60 Hz Only use 552xA
18 C-17 100 mA, 60 Hz Only use 552xA
19 C-18 300 mA, 60 Hz Only use 552xA
20 C-19 300 mA, 60 Hz Only use 552xA
21 C-20 0 A, 0 Hz Use 552xA and 50 turns coil
22 C-21 6 mA, 60 Hz Use 552xA and 50 turns coil
23 C-22 20 mA, 60 Hz Use 552xA and 50 turns coil
24 C-23 60 mA, 60 Hz Use 552xA and 50 turns coil
25 C-24 60 mA, 60 Hz Use 552xA and 50 turns coil
26 C-25 0 A, 0 Hz Use 552xA and 50 turns coil
27 C-26 60 mA, 60 Hz Use 552xA and 50 turns coil
28 C-27 200 mA, 60 Hz Use 552xA and 50 turns coil
29 C-28 600 mA, 60 Hz Use 552xA and 50 turns coil
30 C-29 600 mA, 60 Hz Use 552xA and 50 turns coil
31 C-30 0 A, 0 Hz Use 552xA and 50 turns coil
32 C-31 600 mA, 60 Hz Use 552xA and 50 turns coil
33 C-32 900 mA, 60 Hz Use 552xA and 50 turns coil
34 C-33 1.2 A, 60Hz Use 552xA and 50 turns coil
35 C-34 1.2 A, 60Hz Use 552xA and 50 turns coil
36 Save Push to save STBY