Mechanical Specifications
Size (L x W x H) .....................................232 mm x 85 mm x 45 mm
Weight.....................................................370 g
Environmental Specifications
Operating Temperature........................... -10 °C to +50 °C
Storage Temp .........................................-40 °C to +60 °C
Operating Humidity .................................Non condensing (< 10 °C)
90 % RH (at 10 °C to 30 °C)
75 % RH (at 30 °C to 40 °C)
45 % RH (at 40 °C to 50 °C)
(Without Condensation)
Operating Altitude ...................................3000 meters
Storage Altitude ...................................... 12,000 meters
EMI, EMC................................................Meets all applicable requirements in EN61326-1:2006
Temperature Coefficients........................Add 0.1 x specified accuracy for each degree C above
28 °C or below 18 °C
Safety Compliance.................................. ANSI/ISA S82.02.01:2004
CAN/CSA-C22.2 No. 61010-1-04
IEC/EN 61010-1:2001, 600V CAT III, 300V CAT IV
Pollution Degree 2
EN/IEC 61010-2-32:2002
CAN/CSA-C22.2 No. 61010-2-032 :04
P
Agency Approvals...................................
), ;, ®
Batteries..................................................2 AA, NEDA 15A, IEC LR6