Clamp Meter
9
Adjustment Procedure
Test
(Switch Position)
Product
LCD
Reading
Action
CAL1 then
0.0 V ac
Calibrator output 0 V @ 50 HZ. Push to measure. Allow the
reading to stabilize. Push again to store the value.
K
AC Volts
200.0 V ac
Calibrator output 200.0 V @ 50 HZ. Push to measure. Allow the
reading to stabilize. Push again to store the value.
L
DC Volts
CAL1 then
200.0 V dc
Calibrator output 200.0 V dc. Push to measure. Allow the
reading to stabilize. Push again to store the value.
CAL1 then
300.0 e
Calibrator output 300 Ω. Push to measure. Allow the reading to
stabilize. Push again to store the value.
Ohms
3000 e
Calibrator output 3000
Ω. Push to measure. Allow the reading to
stabilize. Push again to store the value.
CAL1 then
0.0 A ac
Calibrator output 0.0 A @ 50 HZ. Push to measure. Allow the
reading to stabilize. Push again to store the value.
?
AC Amps
(with 50-turn
coil)
200.0 A ac
Calibrator output 4 A @ 50 HZ. Push to measure. Allow the
reading to stabilize. Push again to store the value.
CAL1 then
40.0 μF
Calibrator output 40.0 μF. Push to measure. Allow the reading to
stabilize. Push again to store the value.
E
Capacitance
400.0 μF
Calibrator output 400.0 μF. Push to measure. Allow the reading
to stabilize. Push again to store the value.