Adjustment Procedure
Test
(Switch Position)
Product
LCD
Reading
Action
CAL1 then
0.0 V ac
Calibrator output 0 V @ 50 HZ. Push to measure. Allow the
reading to stabilize. Push again to store the value.
K
AC Volts
200.0 V ac
Calibrator output 200.0 V @ 50 HZ. Push to measure. Allow the
reading to stabilize. Push again to store the value.
L
DC Volts
CAL1 then
200.0 V dc
Calibrator output 200.0 V dc. Push to measure. Allow the
reading to stabilize. Push again to store the value.
CAL1 then
300.0 e
Calibrator output 300 Ω. Push to measure. Allow the reading to
stabilize. Push again to store the value.
Ohms
3000 e
Calibrator output 3000
Ω. Push to measure. Allow the reading to
stabilize. Push again to store the value.
CAL1 then
0.0 A ac
Calibrator output 0.0 A @ 50 HZ. Push to measure. Allow the
reading to stabilize. Push again to store the value.
?
AC Amps
(with 50-turn
coil)
200.0 A ac
Calibrator output 4 A @ 50 HZ. Push to measure. Allow the
reading to stabilize. Push again to store the value.
CAL1 then
40.0 μF
Calibrator output 40.0 μF. Push to measure. Allow the reading to
stabilize. Push again to store the value.
E
Capacitance
400.0 μF
Calibrator output 400.0 μF. Push to measure. Allow the reading
to stabilize. Push again to store the value.

Other manuals for Fluke 373

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 373 and is the answer not in the manual?

Fluke 373 Specifications

General IconGeneral
Operating altitude0 - 3000 m
Non-operating altitude0 - 12000 m
Storage temperature (T-T)-40 - 60 °C
Operating temperature (T-T)-10 - 50 °C
Operating relative humidity (H-H)0 - 90 %
Product colorBlack, Orange
Ergonomic designYes
Continuity30 Ω
DC voltage600 V
Resistance (max)6000 Ω
Resistance (min)600 Ω
Measuring category-
DC voltage resolution (min)0.1 V
AC current maximum resolution (probe)600 A
Battery typeAA
Battery technologyAlkaline
Number of batteries supported2
Weight and Dimensions IconWeight and Dimensions
Depth85 mm
Width232 mm
Height45 mm
Weight370 g