Specifications
29
Frequency via Jaw
Range
375 FC and 376 FC ........................... 5.0 Hz to 500.0 Hz
Resolution ............................................... 0.1 Hz
Accuracy ................................................. 0.5 % ±5 digits
Trigger Level ........................................... 5 Hz to 10 Hz, 10 A
10 Hz to 100 Hz, 5 A
100 Hz to 500 Hz, 10 A
Frequency via Flexible Current Probe
Range
375 FC and 376 FC ............................. 5.0 Hz to 500.0 Hz
Resolution ............................................... 0.1 Hz
Accuracy ................................................. 0.5 % ±5 digits
Trigger Level ........................................... 5 Hz to 20 Hz, 25 A
20 Hz to 100 Hz, 20 A
100 Hz to 500 Hz, 25 A