Maintenance
13
8. Continue to press e to cycle through:
Empty battery annunciator
25 % battery annunciator
50 % battery annunciator
75 % battery annunciator
Blue LED + live battery annunciator (for FC model only, step is skipped for non-FC
model)
White backlight + live battery annunciator
Green backlight + live battery annunciator
9. Press e again to show CAL on the display.
10. Press e again to loop back to show all LCD segments (step 6).
To exit Maintenance Mode at any time, turn off the Meter.
How to Clean the Case
Wipe the case with a damp cloth and mild detergent.
W Caution
Do not use abrasives, isopropyl alcohol, or solvents to clean the case or
lens/window.
Battery Replacement
XW Warning
To prevent personal injury and for safe operation of the Product:
The battery door must be closed and locked before you operate the
Product.
Remove all probes, test leads, and accessories before the battery door is
opened.
Replace the batteries when the low battery indicator shows to prevent
incorrect measurements.
When batteries are changed, ensure that the calibration seal in the battery
compartment is not damaged. If damaged, the Product may not be safe to
use. Return the Product to Fluke for replacement of the seal.