Calibration Manual
2
Performance Test Procedures
Table 1 is a list of the standard equipment for the Performance Tests.
All procedures refer to the performance tests and the readout limits in Table 2.
AC Voltage
To verify ac voltage:
1. Connect the normal output HI and LO of the calibrator to the V Ω and COM input
connections on the bottom of the device under test (DUT).
2. Set the DUT to V (AC Volts measurement mode).
3. Set the calibrator NORMAL output to the voltage and frequency.
4. Apply the voltage to the DUT:
a. Verify that the DUT reads within the display limits.
b. Verify the DUT readings for the remaining steps for AC voltage.
Table 1. Standard Equipment
Equipment Required Characteristics Recommended Model
Calibrator 4.5-digit resolution Fluke 55xxA
Transconductance Amplifier 100 A Fluke 52120A
FieldSense Test Fixture NA PN 2041567
Wired Coil 25-turn 52120A/3kCOIL
Wired Coil 50-turn 5500A/COIL
Test Lead NA PN 2070140