EasyManua.ls Logo

Fluke 377 - Key Features and Concepts; Radio Frequency Data

Fluke 377
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Clamp Meter
Performance Test Procedures
5
AC Current (iFlex Current with AC Voltage Simulation)
To verify ac current measurement with the iFlex, use the AC voltage simulation from the
calibrator:
1. Connect the normal output HI to the Calibrator Output HI to the iFlex 2 mm jack
flexible probe input terminal and the Calibrator Output LO goes to the black COM
jack . See Figure 1.
Figure 1. iFlex Current Probe Connections
2. Set the DUT to K (iFlex AC current measurement mode).
3. Set the calibrator to the voltage and frequency:
a. Verify that the DUT reads within the display limits.
b. Verify the DUT reading for the remaining steps for iFlex Current Probe (with
simulation).
0
/
+
5522A

Other manuals for Fluke 377

Related product manuals