Introduction and Specifications
Specifications
1
1-5
The following additional accessories are supplied with the Fluke 41B:
RS-232 Cable
9-Pin to 25-Pin Adapter
Plug Adapter
3.5 inch Micro-Floppy Disk (FlukeView™ 41 Software)
Optional accessories for both the Fluke 39 and 41B are as follows:
80i-1000s AC Current Probe
C41s Soft Carrying Case
1-8. Operating Instructions
Operating instructions for the Fluke 39 and 41B can be found in the Users Manual
(Fluke PN 942847). See How to Obtain Parts on page 6-1.
1-9. Specifications
Accuracy is specified for a period of one year after calibration.
Specifications expressed in the "±(2% + 3 digits)" format are saying that the error is a percent of
reading plus a number of least significant digits. The least significant digit is the smallest value that
can be displayed in a range and can be found in the resolution section of the Range and
Resolution specifications table.
When a harmonic error is expressed in the "13
th
(±(2% + 2 digits)) --- 31
st
(±(5% + 2 digits))" form, it
is indicating that there is a linear increase in error from the lower harmonic to the upper harmonic.
The error can be determined as a ratio of the desired harmonic to the number of harmonics in the
expression times the difference in error over the range plus the least error. For example, to find the
22
nd
Harmonic error from the above equation:
((Desired Harmonic - Lower Harmonic) / (Upper Harmonic - Lower Harmonic)) X Error Difference +
Least Error
((22 - 13) / (31 - 13)) X 3 + 2 = 3.5%
Therefore, the error for the 22
nd
harmonic is ±(3.5% + 2 digits).
Frequency Range, Fundamental: 6-65 Hz and dc
Minimum Input Levels: 5V rms or 1A rms
Volts Measurements (True rms):
Input Range: 5.0V to 600V rms (ac + dc)
5.0V to ±933V peak
Basic Accuracy*:
rms (ac + dc): ±(0.5% + 2 digits); 6-65 Hz
peak, dc: ±(2% + 3 digits)
*
< 15V rms, add 2 digits
Input Impedance: 1 M, balanced
Crest Factor: > 3.0 below 300V, 1.56 @ 600V
Amps Measurements (True rms)
(1 mV/A) Isolated Input