Introduction and Specifications
Specifications
1
1-7
Other Measurement Specifications
Measurement Function Range/Resolution Accuracy
Input Bandwidth: (-0.5 dB) DC, 6 Hz to 2.1 kHz
Crest Factor (CF): (Using Smooth
%20):
1.00 to 5.00 ±4%
Power Factor (PF): 0.00 to 1.00 ±0.02
Power Factor Displacement (PFD): 0.00 to 0.29
0.30 to 0.69
0.70 to 0.89
0.90 to 1.00
Unspecified
±0.04
±0.03
±0.02
Phase Measurement Range: -179 to 180 degrees (see Phase above)
K-Factor (KF) Range (Model 41B): 1.0 to 30.0 ±10%
Total Harmonic Distortion:
%THD-F:
%THD-R:
0.0 to 799.9
0.0 to 99.9
±(0.03 × Reading + 2.0%)
±(0.03 × Reading + 2.0%)
Ranges and Resolution
AC Volts AC Amps Watts
Range
(PK)
Resolution
Range
(PK)
Resolution
Range
(PK)
Resolution
20V 0.1V 2A 0.01A 50W 1.0W
50V 0.1V 5A 0.01A 100W 1.0W
100V 0.1V 10A 0.01A 200W 1.0W
200V 0.1V 20A 0.01A 500W 1.0W
500V 1V 50A 0.1A 1 kW 0.01 kW
1 kV 1V 100A 0.1A 2 kW 0.01 kW
200A 0.1A 5 kW 0.01 kW
500A 1A 10 kW 0.1 kW
1000A 1A 20 kW 0.1 kW
2000A 1A 50 kW 0.1 kW
100 kW 1 kW
200 kW 1 kW
500 kW 1 kW
1 kkW 1 kW
2 kkW 1 kW