Performance Testing and Calibration
Performance Tests
4
4-7
7. Press f until AP is displayed in the top status line of the Tester.
8. Set the output of calibrator "A" to the Calibrator Output values in Table 4-3. DC
voltages are listed in the upper section of the table and AC voltages at 50 or 60 Hz,
whichever is your local line frequency, are listed in the lower section of the table.
Blank fields do not need to be evaluated.
Note
Apply the voltage incrementally, so the Tester does not autorange to the
next higher range. Ensure the Tester is in the appropriate range before
checking the reading against the performance limits.
9. Verify the readings are within the minimum and maximum limits specified in
Table 4-3.
Note
It is normal for the tester to display “**OL*%THD*” with a VDC input.
Table 4-3. Amps Performance
Calibrator
Performance Limits
Output Fluke 39/41B A rms A pk A dc
mV dc Range Min Max Min Max Min Max
1.00 mV 2 0.94 1.06
6.70 mV 10 6.52 6.88
13.50 mV 20 13.19 13.81
67.00 mV 100 65.2 68.8
135.0 mV 200 131.9 138.1
670.0 mV 1000 652 688
1000.0 mV 2000 976 1024
mV ac
1.00 mV 2 0.96 1.04 1.34 1.49
6.70 mV 10 6.63 6.77 9.25 9.71
13.5 mV 20 13.40 13.60 18.67 19.52
67.00 mV 100 66.3 67.7 92.5 97.1
135.00 mV 200 134.0 136.0 186.7 195.2
670.0 mV 1000 663 677 925 971
1000.0 mV 2000 992 1008 1381 1447