EasyManua.ls Logo

Fluke 39 - Page 74

Fluke 39
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39/41B
Service Manual
5-14
SCK* line (U22-2) should be changing states at slightly less than a 500 KHz rate.
Data written to the serial EEPROM (U22-3) should be the same as U2-33 (line
labeled DIN).
Data read from the serial EEPROM (U22-4) should be the same as the data supplied
to the processor (U2-38).
If the data on U22-4 and U2-38 are not the same, but the ATOD_EE (U3-12) is low,
U3 may be defective.

Table of Contents

Other manuals for Fluke 39

Related product manuals