EasyManua.ls Logo

Fluke 572 - Specification of the Thermometers

Fluke 572
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
572/574/576
Specification of the Thermometers
Temp. Range - 30 to 900°C (- 25 to 1600°F)
Display Resolution 0.1°C (0.2°F)
Accuracy (Infrared)
at 25°C (77°F)
ambient temperature
± 0.75% of reading or
± 0.75°C (± 1.5°F), whichever is
greater ± 2°C (± 4°F) for targets
below -5°C (23°F)
Ambient Derating < 0.05K/K or < 0.05%/K,
whichever is greater at
+ 25°C (77°F) ± 25° (± 45°F)
Optical Resolution
(Standard Focus)
60:1 (19mm spot size at 1.15 m)
(0.75in. spot size at 3.8 feet)
Optical Resolution
(Close Focus)
50:1 ( 6mm spot size at 0.3 m)
(0.24in. spot size at 0.98 feet)
Accuracy
(Thermocouple K & J)
± 2°C or ± 0.75%,
whichever is greater
Accuracy (Thermistor) -30 to 0°C (-22 to 32°F) ± 0.6K
0 to 70°C (32 to 158°F) ± 0.4K
70 to 100°C (158 to 212°F) ±1K
100 to 120°C (212 to 248°F) ±1.5K
Repeatability
(Infrared)
± 0.5% of reading or ± 0.5°C (1°F),
whichever is greater, ± 1°C (± 2°F)
for targets below -5°C (23°F)
Response Time (95%) 250 mSec
Hot Spot Detection (30%) 85 mSec
Spectral Range 8 to 14 µm
Ambient Operating Range 0 to 50°C (32 to 122°F)
Storage Temperature -20 to 50°C (-4 to 122°F) without
batteries
Relative Humidity 10 to 90% at 30°C (86°F),
non condensing
Analog Output
(optional cable needed)
1 mV/°C (°F)
Digital Output RSB232 (Model 574)
USB 1.1 (Model 576)
Power 2 x 1.5 V Alkaline Type AA
Dimensions 240 x 170 x 50 mm
(7.9 x 6.7 x 2 inches)
Tripod Mount 1/4”-20 UNC

Other manuals for Fluke 572