5730A Sales and Demo Guide
Fluke Corporation Company Confidential Page 25 of 28
23 What version MET/CAL is required to run?
Met/CAL V8.2 is required to utilize all 5730A capabilities and
functions.
24
Will the 5730A run with 57XX MET/CAL
procedures?
Yes, the 5730 will run with existing MET/CAL procedures (by
setting the unit to emulate 5700A or 5720A either through the
front panel or remotely).
When using the new features (and performance
specifications) of the 5730A, MET/CAL needs to be upgraded
to V8.2 or newer and procedures will need to be modified to
utilize the new features.
25 Is there a NI Lab view driver for the 5730A? No, not at the moment.
26
Do I need an IEEE card to control the 5730A
with my PC?
No, the 5730A has four industry accepted interfaces: IEEE,
Ethernet, RS-232 and USB. IEEE is required for use with
Service
27
Where do I send back my 5730A for
calibration?
To a Fluke Calibration Service Center. All options are available
online at www.flukecal.eu. However, if using Artifact
Calibration, Fluke Calibration recommends sending the unit in
every two years for a full verification. Normal calibration
28
Can I get a Gold / Silver Care Plan for my
5730A?
Yes. All standard Gold/Silver Care Plans are available for he
5730A.
29
Do you still offer a Gold / Silver Care Plan for
my 5700A and 5720A?
Yes, 1 Yr and 3 Yr CarePlans will be available, but length of
Care plan will be determined by period until obsolescence
date of 5700A / 5720A calibrator.
30
What manuals, data, information come with
the 5730A?
The unit ships with a printed Getting Started Guide in 1 of 9
languages. The CD includes the full Operators Manual in
English and the Calibration Manual in English. It also includes
the required USB driver. Chinese, Japanese, and Russian
Operators Manuals will be coming soon. There will also be a
DVD included which contains the Linux Source Code for the
31 What accessories does the 5730A ship with?
The 5730A ships with an AC power cord, Calibration
Certificate, Operators Manual CD, and Printed Getting Started
Artifact Cal and Cal Check
30
Does the 5730A use the same "Artifact Cal
process" as the 5700A/ 5720A?
Yes, it utilizes the same proven Artifact Cal methodology that
has been used over the last 25 years in the 5700A and 5720A.
31 How does Artifact Cal give me better specs?
Artifact Cal allows you to calibrate your instrument against the
required standards in less than an hour. Through this, the
5730A can be used against 90 day or even 24 hour specs,
depending on how often Artifact Cal is performed.
32
What equipment is needed to artifact calibrate
the 5730A?
The only instruments required to calibrate the 5730A are a
732B , 742A-1, 742A-10K plus one of the recommended 5730A
Lead kits (5440A-7003 Spade Lugs recommended for Art Cal).