5730A Sales and Demo Guide
Fluke Corporation Company Confidential Page 26 of 28
33
When I perform an Artifact Cal, how will this
affect my scope of accreditation?
If you are accredited using the 1-year specifications for your
scope and then perform the Artificat Cal, you are OK. If you
are accredited using the 90-day specification, then you will be
required to run the Artificat Cal or utilize some other type of
metrology to achieve the 90 day specifications.
34 What is Cal Check?
Cal Check is an internal calibration method which utilizes the
internal reference to verify the performance of the internal
circuitry. This then provides internal drift data to the USB
memory device which may be used for drift/uncertainty
analysis and charting. It does not allow the user to change
any of the calibration constants; it is simply a check.