6200-2
Users Manual
16
Note
Accidental measurement of a defective unit may trip a RCCB (residual
current circuit breaker).
Load/Leakage Current Test
Warning
To prevent possible electrical shock, fire, or personal injury, NEVER
carry out this test unless you have first carried out a thorough visual
inspection, followed by a test of the earthing (Class I appliances), and
then a test of the insulation. You must verify that these tests are passed
before you do this test.
Caution
Live test. The appliance is energized at mains voltage. For this purpose,
switch on the appliance. Appliances driven by motors or equipped with
heating units may present a danger for the person testing (comply with
the appliance instruction manual). Ensure that the appliance is in a safe
condition to operate and secure before a test.
The Load/PE Leakage test consists of:
• a fuse and L-N loop pre-test
• measurements of the appliance power consumption and load current at full mains voltage
• measurement of the earth leakage current (differential measurement) at full mains
voltage.
The measurements are performed in one test sequence.
To perform the Load/PE Leakage test:
1. Connect the appliance and the test lead as indicated on the Tester, see Figure 5.
MEM
PRINT
BOND
200 mA
BOND
25 A
INSUL-
ATION
I
SUB
LOAD/
LEAK
I
TOUCH
IEC
LEAD
ZERO
CLEAR
I
P
E
<
3
.5
m
A
(
0.
75mA
P
OR
T
AB
L
E
A
P
P
LIANC
E
)
I
N
S
U
L
AT
I
O
N
I
S
U
B
B
O
ND ZE
R
O
I
E
C
L
E
A
D
EARTH BOND
L
OAD
/
L
E
AK
I
T
O
UC
H
I
T
O
U
C
H
<
0
.
25
m
A
R
I
S
O
>
1
.0
M
C
l
a
s
s
I
R
I
S
O
>
2
.0
M
C
l
a
s
s
I
I
RPE
<
0
.
1
R
I
S
O
>
1
.
0
M
C
l
a
s
s
I
R
PE
20
0
mA
,
R
P
E
2
5
A
R
PE
<
0.1
+
L
E
A
D
P
R
O
B
E
O
N
L
Y
WI
TH
C
LASS
II
A
PP
L
I
AN
C
E
6200-2
A
PPLIANCE TESTE
R
hdc05.eps
Figure 5. Load/Leakage Connections