Appliance Tester
Specifications
25
Test Limits for PASS result
UK AU DE NL
Earth Bond 200mA <0.10 Ω <1.0 Ω <0.30 Ω <0.30 Ω
Earth Bond 25A <0.10 Ω <1.0 Ω NA <0.30 Ω
Earth Bond 10A NA NA <0.30 Ω NA
Insulation Class I >1 MΩ >1 MΩ >1 MΩ >1 MΩ
Insulation Class II >2 MΩ >1 MΩ >2 MΩ >2 MΩ
Substitute Leakage Class I <3.5 mA <1.0 mA <1.0 mA <1.0 mA
Substitute Leakage Class II <0.50 mA <1.0 mA <0.50 mA <0.50 mA
Leakage Current <0.75 mA <5.0 mA <3.5 mA <3.5 mA
Touch Current >0.25 mA >1.0 mA <0.50 mA <0.50 mA
IEC lead Earth Bond <0.10 Ω <1.0 Ω <0.30 Ω <0.30 Ω
IEC lead Insulation >2 MΩ >1 MΩ >1 MΩ >1 MΩ
Variation Factor Errors
Variation Factor Designation % Variation Error
Position E1 0.0 %
Supply Voltage E2 5.0 %
Temperature E3 5.5 %
Current Consumption E4 1.5 %
Magnetic Fields E5 2.5 %
Impedance E6 1.0 %
Capacitance E7 2.0 %
Current Waveshape E8 1.0 %