22
Scale mA Channel to Temperature
The Product provides a function to convert mA current
channel reading to temperature reading.
To scale mA channel to temperature:
1. Push .
2. Push .
The temperature reading shows on the display.
To switch back to mA channel:
1. Push .
2. Push .
The mA current reading shows on the display.
Note
Temperature at 4 mA = Span check 0 %
Temperature at 20 mA = Span check 100 %
Set 0 % and 100 % Output Parameters
You must set the 0 % and 100 % source values before
you can use the step and ramp functions. Proceed as
follows:
1. If necessary, push for Source mode.
2. Use the arrow keys to enter the value for 0 %.
3. Push and then to set the 0 % value.
4. Use the arrow keys to enter the value for 100 %.
5. Push and then to set the 100 % value.
6. Use , , , or to adjust the value.
Note
This function is only available when Source
mode is selected.